BAXTER HEALTHCARE CORPORATION FLO-THRU; CLAMP, VASCULAR Back to Search Results
Catalog Number FT12225
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/23/2022
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a hair-like object was attached to the tab of a 2.25mm flo-thru device.This was observed during an incoming inspection.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the device was received for evaluation.Visual inspection along with magnification was performed which observed a hair-like object statically clung to the flo-thru tab which was identified to be most likely an eyelash or a small piece of hair.The size of the hair or eyelash that was identified did not exceed the specification limits.The condition was identified; however, the product met specifications.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLO-THRU
Type of Device
CLAMP, VASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
SYNOVIS SURGICAL INNOVATIONS
2575 university ave. w
saint paul MN 55114
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15048035
MDR Text Key297086755
Report Number1416980-2022-03693
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K981624
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberFT12225
Device Lot NumberSP22D051652162
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received07/21/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-