Catalog Number FT12225 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/23/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that a hair-like object was attached to the tab of a 2.25mm flo-thru device.This was observed during an incoming inspection.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation.Visual inspection along with magnification was performed which observed a hair-like object statically clung to the flo-thru tab which was identified to be most likely an eyelash or a small piece of hair.The size of the hair or eyelash that was identified did not exceed the specification limits.The condition was identified; however, the product met specifications.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|