• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 15750 ALTON PKWY RAD-67 SPOT CHECK PULSE CO-OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 15750 ALTON PKWY RAD-67 SPOT CHECK PULSE CO-OXIMETER Back to Search Results
Model Number 27692
Medical Device Problem Code High Readings (2459)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/25/2022
Type of Reportable Event Malfunction
Event or Problem Description
The customer reported inaccurate high sphb values compared to lab results.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
The product has been returned to masimo for evaluation.When the investigation is complete a follow up report will be submitted.
 
Additional Manufacturer Narrative
Additional manufacturing narrative: the returned device was evaluated.During evaluation the device passed all visual and functional testing.During simulation testing, the device passed manual and preset conditions and provided accurate measurements.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.
 
Event or Problem Description
The customer reported inaccurate high sphb values compared to lab results.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-67 SPOT CHECK PULSE CO-OXIMETER
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 15750 ALTON PKWY
15750 alton pkwy
irvine CA 92618
Manufacturer (Section G)
MASIMO - MEXICALI
industrial vallera de mexicali calzada del oro, no.2001
mexicali, baja california 21600
MX   21600
Manufacturer Contact
matofa fautua
15750 alton pkwy
irvine, CA 92618
9498223661
MDR Report Key15048286
Report Number3019388613-2022-00157
Device Sequence Number5439133
Product Code DQA
UDI-Device Identifier00843997012812
UDI-Public00843997012812
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K040214
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 06/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number27692
Device Catalogue Number9794
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/08/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/21/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date07/18/2022
Supplement Report FDA Received Date08/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/18/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DCI-MINI SC 400; M20-01 PATIENT CABLE
Patient Age25 YR
Patient SexFemale
Patient Weight63 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-