• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR; DH CPK NG TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR; DH CPK NG TUBES Back to Search Results
Model Number 40-7368
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
Avanos medical inc.Received a single report that referenced two different incidences, which were associated with separate units, involving two different patients.This is the second of two reports.Refer to 9611594-2022-00100 for the first report.The guide of the new catheter was tested before placing it and no problem was encountered; however, once the catheter was in place and the guide was removed a crack was noted, that was reportedly inside the patient.The newly placed tube was removed; there was no report of injury.
 
Additional Manufacturer Narrative
The actual complaint product was not returned for evaluation.The device history record for lot 30151280 was reviewed and the product was produced according to product specifications.Root cause could not be determined.All information reasonably known as of 18 jul 2022 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORFLO NASOGASTRIC/NASOINTESTINAL FEEDING TUBE WITH STYLET WITH ENFIT CONNECTOR
Common Device Name
DH CPK NG TUBES
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
Manufacturer (Section G)
AVENT SA DE R.L. DE C.V. (AVENT 1)
circuito industial no.40
colonia obrera
nogales, cp 84048
MX   84048
Manufacturer Contact
lisa clark
5405 windward parkway
alpharetta, GA 30004
4704485444
MDR Report Key15048291
Report Number9611594-2022-00101
Device Sequence Number5439136
Product Code FPD
UDI-Device Identifier00350770460314
UDI-Public00350770460314
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K083210
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/18/2026
Device Model Number40-7368
Device Catalogue NumberN/A
Device Lot Number30151280
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2022
Initial Report FDA Received Date07/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/27/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age4 YR
Patient SexMale
Patient Weight14 KG
-
-