• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC DAVINCI XI; PATIENT SIDE CART, 4-ARM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC DAVINCI XI; PATIENT SIDE CART, 4-ARM Back to Search Results
Model Number 380652-37
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/20/2022
Event Type  malfunction  
Manufacturer Narrative
An intuitive surgical, inc.(isi) field service engineer (fse) was dispatched to the customer site to further investigate the reported complaint.The fse replaced the universal surgical manipulator (usm).The system was tested and verified as ready for use.Isi received the part(s) involved with this complaint and completed the device evaluation.Failure analysis investigation confirmed and replicated the reported event.The unit was tested on an in-house system and passed normal mode.The unit failed a pftp as it failed the pitch friction very slow and speed low test.The unit went through more testing on a pftp and passed direction tests, lissajous, cva characterization, sine cycle, carriage friction test, brake release test, brake hold test, advanced brake test, carriage strength test, and carriage switches test.The pitch motor module, pitch harmonic drive, and elbow bearings will be replaced as the fix the reported problem.A log review confirmed the procedure date of (b)(6) 2022 on system sk1135.No image or procedure video was provided for review.Based on the information available at this time, this complaint is being classified as a reportable event due to the following conclusion: a usm was abandoned after the start of the procedure and the surgeon was able to continue with the procedure robotically using 3 usms.While there was no harm or injury to the patient, the reported failure mode could likely cause or contribute to an adverse event if it were to recur.System unavailability after start of a surgical procedure could lead to the procedure to be converted/aborted and may lead to an injury due to the patient's inability to tolerate a conversion/abortion.
 
Event Description
It was reported that during a da vinci-assisted incisional hernia surgical procedure, a customer was having an ongoing issue with an instrument on universal surgical manipulator 3 (usm3).Technical support was contacted and the technical support engineer (tse) was informed that the field service engineer (fse) was recently onsite however the issue remains.Prior to calling tech support, the customer had switched from usm3 and is using usm4 for a 3 armed procedure.The tse requested an image of usm3 from the customer which was provided after the call ended.The tse noted the flex strips are being shaped opposite of the guidance provided in the da vinci xi user manual.Live logs were reviewed and no associated issues were identified.The procedure was completing as planned with no reported injury.Intuitive surgical, inc.(isi) made multiple follow-up attempts to obtain additional information.However, no further details have been received as of the date of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DAVINCI XI
Type of Device
PATIENT SIDE CART, 4-ARM
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 
4085232100
MDR Report Key15048671
MDR Text Key304533671
Report Number2955842-2022-13036
Device Sequence Number1
Product Code NAY
UDI-Device Identifier00886874110720
UDI-Public(01)00886874110720
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131861
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Nurse
Remedial Action Other
Type of Report Initial
Report Date 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number380652-37
Device Catalogue Number380652
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/24/2022
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/20/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
DA VINCI INSTRUMENTS AND ACCESSORIES
-
-