MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH C-TAPER COCR LFIT HEAD 32MM/+5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH C-TAPER COCR LFIT HEAD 32MM/+5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number 06-3205
Medical Device Problem Codes Corroded (1131); Degraded (1153)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Cancer (3262)
Date of Event 06/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a primary total hip arthroplasty with surefit stem and 32 mm +5 head and trident cup.The patient later presented with a painful right hip and required a revision due to possible trunnionosis and possible pseudotumor.Initial reporter confirmed trunnionosis was present & pseudotumor was sent to pathology for analysis.
 
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Device not returned to the manufacturer.
 
Additional Manufacturer Narrative
Reported event: an event regarding altr and fretting/trunnionosis involving a metal head was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection of the returned device indicates nothing remarkable.No visible black debris.Minor scratches and dents observed consistent with explantation.Clinician review: insufficient information was provided to support a medical review.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event could not be determined because insufficient information was provided.Further information such as return of the device, histopathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that the patient underwent a primary total hip arthroplasty with surefit stem and 32 mm +5 head and trident cup.The patient later presented with a painful right hip and required a revision due to possible trunnionosis and possible pseudotumor.Initial reporter confirmed trunnionosis was present & pseudotumor was sent to pathology for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
C-TAPER COCR LFIT HEAD 32MM/+5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15050781
Report Number0002249697-2022-01034
Device Sequence Number17348829
Product Code JDI
UDI-Device Identifier07613327016321
UDI-Public07613327016321
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K910988
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2007
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/29/2012
Device Model Number06-3205
Device Catalogue Number06-3205
Device Lot Number809MDDX1
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/27/2007
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age78 YR
Patient SexFemale
Patient Weight71 KG
-
-