• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 HICKMAN CV CATHETER, DUAL-LUMEN, 9F; CHRONIC CATHETERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 HICKMAN CV CATHETER, DUAL-LUMEN, 9F; CHRONIC CATHETERS Back to Search Results
Model Number 0600600
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiration date: 02/2026).
 
Event or Problem Description
It was reported that two days post chronic catheter placement, the catheter allegedly fractured.The procedure was completed using another device.There was no reported patient injury.
 
Event or Problem Description
It was reported that two days post chronic catheter placement, the catheter allegedly fractured.The procedure was completed using another device.There was no reported patient injury.
 
Additional Manufacturer Narrative
Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one 9fr hickman d/l catheter was returned for evaluation.Functional, gross visual, tactile, and microscopic visual evaluations were performed.A complete circumferential break was noted on the distal end of the catheter.However, striations were noted throughout.The investigation is confirmed for the reported fracture issue, as a split was noted on the white luer extension leg.Upon infusion of the white luer, a leak from the split was noted.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiration date: 02/2026).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HICKMAN CV CATHETER, DUAL-LUMEN, 9F
Common Device Name
CHRONIC CATHETERS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15050906
Report Number3006260740-2022-02761
Device Sequence Number4667890
Product Code LJS
UDI-Device Identifier00801741051890
UDI-Public(01)00801741051890
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K830233
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0600600
Device Catalogue Number0600600
Device Lot NumberHUFW2622
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer09/23/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
-
-