As the lot number for the device was provided, a review of the device history records is currently being performed.The device has not been returned to the manufacturer for evaluation.However, photos and images were provided for review.The investigation of the reported event is currently underway.(expiration date: 11/2023).
|
|
The manufacturing location was selected as unknown due to system limitations.The manufacturing location for the encor product is adroit medical equipment.Manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one encor biopsy probe was returned for evaluation.On visual evaluation, the device appeared to have residue throughout and the aperture was received in a completely open condition.And it was noted that the distal end of the needle was deformed (i.E twisted/bent).Due to the condition of the device, functional testing was not performed.Therefore, the investigation for the identified material deformation is confirmed as the needle was noted to be deformed.However, the investigation for the reported difficult to remove and the material twisted/bent is kept as inconclusive as the functional testing was not performed.Three electronic photos were provided for review.First photo shows the label where the details were verified with the updated information.Second and third photo shows the needle of the encor probe.And it shows the aperture of the needle is slightly narrowed and it seems like the needle is deformed.Therefore, based on the photo review, the identified failure material deformation can be confirmed as the deformed needle is noted.However, based on the photo review, the reported failure difficult to remove could not be confirmed.Therefore, the investigation for the identified material deformation is confirmed as there were proper evident provided.However, the investigation for the reported difficult to remove and material twisted/bent is kept as inconclusive as there were no proper evident provided.Five electronic images were reviewed.First two image shows the image from the stereotactic biopsy probe within the breast.And it shows the probe is bent near its distal end.Another set of images were noted to be taken after the procedure which shows the slightly deformed needle and it was noted that the chamber is mildly narrowed.Last photograph is the photograph of the packaging of the needle.Based on the image review, the identified material deformation can be confirmed as the deformation was noted.Based on the image review, the reported difficult to remove and material twisted/bent could not be confirmed.A definitive root cause for the identified material deformation ,alleged difficult to remove and alleged material twisted/bent could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiration date: 11/2023).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|