U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752202
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported broken phacoemulsification tip broke out of patient's eye during sculpting phase of the cataract procedure; however, there was no patient contact when this was observed.The broken tip was replaced by a mini tip for completing the procedure.There was no patient harm.There were two products reported from the site as potentially involved the reporter could not provide clarification regarding which product is associated with this complaint.No additional information is available.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
A sample was not received at the manufacturing site for evaluation.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The report of phaco tip broken is confirmed based on the photos attached to parent complaint, however, because a sample was not received and the product was processed and released according to the product¿s acceptance criteria, the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown; therefore, specific action with regards to this complaint cannot be taken.All phaco tips are 100% visually inspected by trained operators using 30x magnification during the manufacturing process.Any nonconformance, such as the broken phaco tip exhibited in the photos attached to the parent complaint, is removed from the lot and scrapped.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15051053
Report Number1644019-2022-00577
Device Sequence Number20177679
Product Code LRO
UDI-Device Identifier00380657522026
UDI-Public00380657522026
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Catalogue Number8065752202
Device Lot Number2408591H
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer09/21/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date10/17/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CUSTOM-PAK SURGICAL PROCEDURE PACK
Patient SexUnknown
-
-