Model Number V-GO 20
Medical Device Problem Code Activation, Positioning or Separation Problem (2906)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Malfunction
Event or Problem Description
A patient reported that over the past 3 or 4 days, they have had three v-go 20 devices where the needle button popped out within 4 minutes of initial needle insertion.The patient confirmed that they press on the raised circular bump of the needle button during initial needle insertion.It was reported that the devices are available for return to the manufacturer for evaluation.
 
Additional Manufacturer Narrative
Device evaluation: three devices were received and evaluated for the complaint regarding the needle button released event.All devices were received, and the needle mechanism functions were tested and verified to have operated as intended.The complaint could not be confirmed for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V-GO DISPOSABLE INSULIN DEVICE
Common Device Name
V-GO
Manufacturer (Section D)
MANNKIND CORPORATION - V-GO
293 boston post road w
#330
marlborough MA 01752
Manufacturer (Section G)
MANNKIND CORPORATION - V-GO
293 boston post road w
#330
marlborough MA 01752
Manufacturer Contact
jeffrey zajac
293 boston post road w
#330
marlborough, MA 01752
MDR Report Key15051567
Report Number1226572-2022-00091
Device Sequence Number14648654
Product Code LZG
UDI-Device Identifier00385609600036
UDI-Public(01)00385609600036(17)230707(10)FG220221
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K103825
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date07/07/2023
Device Model NumberV-GO 20
Device Catalogue Number8560-9600-03
Device Lot NumberVG220122B
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/11/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2022
Initial Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
-
-