MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL GALAXY G3 XSFT 2MM X 2CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL GALAXY G3 XSFT 2MM X 2CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number GLX120202
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Initial reporter: the customer contact information, including name, occupation, phone, fax, and e-mail address, was not reported.The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
As reported by the field, during a coil embolization, a 2mm x 2cm galaxy g3 xsft coil (glx120202, 30504545) was being used as the fourth coil.When the physician tried to detach the coil, the beeping sound was normal but during retrieving delivery system, it continuously came out along the delivery system.The physician used a same like product.The procedure was successfully finished.There was no patient injury reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by the field, during a coil embolization, a 2mm x 2cm galaxy g3 xsft coil (glx120202, 30504545) was being used as the fourth coil.When the physician tried to detach the coil, the beeping sound was normal but during retrieving delivery system, it continuously came out along the delivery system.The physician used a same like product.The procedure was successfully finished.There was no patient injury reported.No additional information is available.A non-sterile galaxy g3 xsft 2mm x 2cm was received contained in the decontamination pouch.Upon receiving the device, a visual inspection was performed, and no apparent damage was observed.The device was inspected under magnification, and it was found that the coil was attached to the unit and is in good normal conditions (i.E., no elongations, kinks, or non-concentric loops were noted).The distal outer sheath had not been softened, indicating that the resistance heating (rh) coil had not been heated and the detachment process was not initiated.The electrical resistance of the galaxy g3 mini 2mm x 3cm device was measured with a multimeter which is within the specification range between 48.5ohms o ¿ 56.0 ohms o.Also, the device was connected to a lab sample detachment control box (b)(4) (dcb) and a lab sample enpower control cable and the power was turned on.The system ready light was illuminating.Then, the detach button was pressed, and a beep sound was heard.A manufacturing record evaluation was performed for the finished device 30504545 number, and no non-conformances related to the malfunction were identified.The customer complaint regarding a failure to detach was not confirmed since the resistance heating (rh) was heated and the coil was successfully detached from the unit.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.A manufacturing record evaluation was performed, and no non-conformances related to the reported complaint condition were identified.Failure to detach is a known potential issue associated with the use of this device.The instructions for use (ifu) contains several precautions related to this issue and includes instructions for troubleshooting the situation should it be encountered during use.It should be noted that product failure could be caused by multiple factors.The instructions for use (ifu) do contain the following recommendations: ¿ verify that the microcoil delivery system is fully connected, and no faults are indicated on the dcb.If a fault exists, reseat all connections between the dpu, the dcb, and the connecting cable.If a fault still persists, replace the connecting cable.If this does not correct the error, replace the dcb.If the microcoil delivery system still continues to have a fault, retrieve the microcoil, re sheathing the microcoil system, and replace it with a new microcoil system.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and the evidence presented by the returned device; however, it is possible that clinical and procedural factors, including device manipulation and device interaction, may have contributed to the reported failures and damages on the returned system.Multiple attempts to obtain additional information were unsuccessful.If information is provided at a later date, the file will be reopened and processed accordingly.As part of the cerenovus quality process, all devices are manufactured, inspected, and released to approved specifications.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the event was related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALAXY G3 XSFT 2MM X 2CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
Manufacturer (Section G)
CODMAN & SHURTLEFF, INC. (FREMONT)
47709 fremont blvd
fremont CA 94538
Manufacturer Contact
michelleann garcia
31 technology dr
irvine, CA 92618
646591-7981
MDR Report Key15051582
Report Number3008114965-2022-00494
Device Sequence Number4784517
Product Code KRD
UDI-Device Identifier10886704077640
UDI-Public10886704077640
Combination Product (Y/N)N
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model NumberGLX120202
Device Catalogue NumberGLX120202
Device Lot Number30504545
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-