DEPUY ORTHOPAEDICS INC US PINN MAR NEUT 32IDX56OD; PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1219-32-056
Medical Device Problem Codes Device Dislodged or Dislocated (2923); Naturally Worn (2988)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/01/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Patient has polyethylene wear outside the regular standards.First surgery performed 4 years ago second surgery performed 2 years ago third surgery will be on july 4.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a manufacturing records evaluation (mre) was not performed as no lot number was provided for this device.H10 additional narrative: added: a2 (age) and a3.
 
Additional Manufacturer Narrative
Product complaint # (b)(6) investigation summary: the device associated with this report was received for analysis.Based in the visual examination of the part, it was observed that a portion of the liner's rim was fractured.The fracture is consistent with excessive edge loading of the femoral head and patient's body weight since machining lines are yet visible at the internal surface.Interlock feature edge of the liner denoted four ard tabs sheared off from the liner which is also evidence of the liner having disassociated from the cup.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : a manufacturing records evaluation (mre) was not performed since a valid finished good lot number was not possible to be retrieved from the part due to damage on the device.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d10 and g4.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: d1, d2 and d4.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3, d10 (concomitants).
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that it was found during the visual examination performed on the involved poly liner, the cup and liner went through a disassociation event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN MAR NEUT 32IDX56OD
Common Device Name
PINNACLE HIP SYSTEM : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15051731
Report Number1818910-2022-13370
Device Sequence Number17477432
Product Code LPH
UDI-Device Identifier10603295013921
UDI-Public10603295013921
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K001534
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1219-32-056
Device Catalogue Number121932056
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer08/02/2022
09/27/2022
10/31/2022
01/11/2023
01/19/2023
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date08/05/2022
09/28/2022
11/01/2022
01/17/2023
01/19/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 32 +1 GR; CORAIL AMT COLLAR SIZE 11; CORAIL AMT COLLAR SIZE 11; PINN MAR NEUT 32IDX56OD; UNK HIP ACETABULAR CUP METAL
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient SexFemale
-
-