During an ankle exam on the exam bed.The equipment was powered off, and the tube was suspended and pressed down on the patient.The patient believed that the ankle fracture was caused by the pressure of our philips equipment.The patient thinks that our philips equipment was pressed down and caused the ankle fracture.Now the hospital asks: if the patient cannot accept the hospital's explanation, the hospital needs philips to cooperate with the explanation.Patient believes he has been harmed.Sudden power failure during normal operation.
|
|
The issue reported was that during an ankle exam, the equipment was powered off, and the tube was suspended and pressed down on the patient and caused the patient's ankle fracture.The philips service engineer inspected system and confirmed the ceiling suspension (cs) counterweight was unbalanced.After counterweight adjustment the device returned to normal use.Philips research and development (r&d) team investigated this issue, and concluded that from product design to product production, product installation and later after-sales service, the spring balancer performance is under control.The device was installed at the hospital back on (b)(6) 2017 and is out of service warranty.The philips equipment record system indicated that the last preventive maintenance (pm) of ceiling suspension of this device was done on (b)(6) 2020, there's no other preventive maintenance performed ever since as acknowledged by the customer.In the ifu (dmr359398) it is stated that digitaldiagnost c50 requires periodic maintenance and philips recommend: the x-ray unit is serviced by the philips customer service at least once a year, must have heavily used x-ray equipment serviced more frequently to avoid endangering the patient and meet obligations.Also, device requires proper operation, planned maintenance, and checks the user must perform routinely, which are essential to keep the product operating safely, effectively and reliably.And the obligations of the user including proper operation, regular testing by the user and regular service and repair.By taking these preventive actions user maintain the operability and operational reliability of the system.As the user of an x-ray equipment user are obliged according to accident prevention regulations, the medical products law and other regulations to take such precautions.User must check ceiling suspension for loose or damaged components or accessories daily.If operational defects or other departures from normal operational behavior occur, must turn off the x-ray unit and inform customer service, may only resume operation of the x-ray equipment when it has been repaired.Operation using faulty components may lead to an increased safety risk or unnecessarily.Philips healthcare products are designed to meet stringent safety standards, and, the system shall not be used, if any part of the equipment or system is known (or suspected) to be operating improperly.Users should follow the safety warnings and guidelines described in the instructions for use (ifu), qrg (quick reference guide), and trg (technical reference guide).It is vital that the users strictly follow all safety directions to ensure the safety of both patients and operators.The probable cause was unbalanced ceiling suspension counterweight, due to the spring balancer wasn't adjusted on a recommended basis.Internal cross reference: complaint (b)(4).
|