MAUDE Adverse Event Report: BIOMET UK LTD. OXFORD PKS CEMENTED TIBIA, ALPHA, STANDARD - RM SIZE D; KNEE PROSTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXFORD PKS CEMENTED TIBIA, ALPHA, STANDARD - RM SIZE D; KNEE PROSTHESIS Back to Search Results
Model Number N/A
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Pain (1994); Loss of Range of Motion (2032); Swelling/ Edema (4577)
Date of Event 04/30/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a revision procedure of the right knee approximately three (3) months post implantation due to pain, swelling and limitation of movement after extreme flexion of the knee.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Additional Manufacturer Narrative
(b)(4).Medical devices: oxf anat brg rt x-sm 4mm pma item# 160791 lot# 6741836; oxford uni twin-peg femoral xs item# 166940 lot# 6839631.Report source: foreign - china.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 3002806535 - 2022 ¿ 00319; 3002806535 - 2022 ¿ 00320.
 
Additional Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.The following sections were updated: b4, b5, g3, g6, h1, h2, h6, h10.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Devices are used for treatment.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: overall fit and alignment are maintained, there is no sign of loosening.No dislocation is identified on the images.However, a dislocation could certainly cause symptoms such as pain and swelling and the need for a revision.The root cause of the reported issue is attributed to patient non compliance based upon the information provided by the surgeon and supported by the operating surgeon's assessment.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXFORD PKS CEMENTED TIBIA, ALPHA, STANDARD - RM SIZE D
Common Device Name
KNEE PROSTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key15052343
Report Number3002806535-2022-00321
Device Sequence Number11955180
Product Code NRA
UDI-Device Identifier05019279388912
UDI-Public(01)05019279388912(17)310112(10)6937758
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P010014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/24/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberN/A
Device Catalogue Number154725
Device Lot Number6937758
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/23/2022
Supplement Date Received by Manufacturer08/02/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
Patient SexFemale
-
-