Medical device problem code (b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot did not indicate a lot specific quality issue.It was reported that the contrast mix used in the procedure was 90% saline/10% contrast.It should be noted that the coronary dilatation catheters (cdc), nc trek rx, global instructions for use (ifu), materials required section, specifies that the contrast is 60% contrast medium diluted 1:1 with normal saline.In this case, since the contrast ratio used was less than the required percentage it is unknown if the ifu violation caused or contributed to the reported complaint.The investigation was unable to determine a conclusive cause for the reported failure to deflate; however, the reported unexpected medical intervention appears to be related to circumstance of the procedure.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
|