ZIMMER ORTHOPAEDIC MFG. LTD. ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 12 MM HEIGHT; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hematoma (1884); Unspecified Infection (1930)
Date of Event 06/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Medical products: natural tibia trabecular metal two-peg porous fixed bearing left size g item# 42530007901 lot# 64912696; femur trabecular metal cruciate retaining (cr) standard porous item# 42502806401 lot# 6498882; all-poly patella cemented 35 mm diameter item# 42540200035 lot# 65271105.Source: australia.Customer has not indicated whether the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that a patient was revised approximately two weeks post implantation due to suspected infection.The patient underwent a washout, and the articular surface was exchanged as an extra cautionary measure.No infection was confirmed, however a hematoma was confirmed.Attempts to obtain additional information have been made; however, no more information is available at this time.
 
Additional Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, g3, g6, h1, h2, h3, h6, and h10 during the investigation process a review of the sterile certifications were reviewed and found to be conforming with no applicable deviations.Devices were verified to have gone through acceptable sterilization process following iso/aami/astm & eu published guidelines.There are multiple factors that may contribute to an infection that are outside the control of zimmer biomet, such as external factors, i.E., hospital/surgical environment, provider related risk factors, and/or patient comorbidities/risk factors.As there are no indications of a product or process issues identified affecting implant safety or effectiveness, therefore implanted products are not identified as the source or contributing to the suspected (unconfirmed) infection.A hematoma is a mass of clotted blood that forms in a tissue, organ, or body space.A hematoma can be associated with pain, swelling, ecchymosis, serosanguinous or bloody drainage after a recent surgical procedure.The development of a postoperative hematoma after a procedure can be correlated with the surgical procedure and perioperative anticoagulation therapy to prevent dvt (prophylaxis).Patients may be at increased risk for developing hematomas if they have received fresh-frozen plasma, vitamin k, or hormonal therapy as well.Most hematomas resolve on their own, without surgical intervention, while some others do not.Larger hematomas may need to be surgically evacuated in order to resolve.As timeframes of onset differ due to individual contributing factors, a specific timeframe of expected occurrence cannot be established.The presence of a hematoma increases risk for infection.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULAR SURFACE MEDIAL CONGRUENT (MC) LEFT 12 MM HEIGHT
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15052797
Report Number3007963827-2022-00188
Device Sequence Number20212554
Product Code MBH
UDI-Device Identifier00889024469396
UDI-Public(01)00889024469396(17)270131(10)65355244
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K150090
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number42512100912
Device Lot Number65355244
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer08/10/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/23/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age50 YR