• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH TI MATRIXRIB PRE-CONTOURED PL 16 HOLES FOR RIGHT RIBS 4 & 5; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH TI MATRIXRIB PRE-CONTOURED PL 16 HOLES FOR RIGHT RIBS 4 & 5; PLATE, FIXATION, BONE Back to Search Results
Model Number 04.501.004
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/21/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Event occurred at an unknown date in 2022 complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from australia reports an event as follows: it was reported that on an unknown date in 2022, the surgeon was using the matrixrib system with the matrix prodriveii to implant a plate and screws on a fractured rib.The plate slipped off the rib slightly after he inserted the first screw.He could not remove the screw as it was lodged in the plate.When he went to insert the second screw to see if he could reposition the plate, it caused a splintering of the rib and was still off position.He said the bone was osteoporotic.He tried to remove the second screw and it too was stuck in the plate.He was able to disengage the plate and screws from the bone with the screws intact in the plate and unable to be removed.He ended up using a competitive system that he was more familiar with.This is not the first time that he has tried to remove the screw once inserted and was unsuccessful.This report is for a ti matrixrib pre-contoured pl 16 holes for right ribs 4 & 5.This is report 2 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.A photo investigation was completed: visual analysis of the photo revealed that there was no damage or defects with the matrixrib plate.There is not enough evidence to confirm the reported allegation based on the pictures provided.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was not confirmed as the matrixrib plate was found to have no damage or defects.No definitive root cause could be determined.There was no indication that a design or manufacturing issue contributed to the complaint.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was noted the procedure was completed on (b)(6) 2022.The surgeon removed the plate from the bone with the screws still lodged in the plate.He then used a non-synthes system.It was noted, fragments were generated but were able to be removed easily without additional intervention.The procedure was completed successfully with a five (5) minute delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI MATRIXRIB PRE-CONTOURED PL 16 HOLES FOR RIGHT RIBS 4 & 5
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key15052945
Report Number8030965-2022-04970
Device Sequence Number5489663
Product Code HRS
UDI-Device Identifier10887587016986
UDI-Public(01)10887587016986
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K081623
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.501.004
Device Catalogue Number04.501.004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/21/2022
Supplement Date Received by Manufacturer08/30/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date09/09/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LOCK SCREW Ø2.9 SELF-TAP L10 TAN; LOCK SCREW Ø2.9 SELF-TAP L10 TAN
Patient Age87 YR
Patient SexMale
-
-