Model Number PS TIBIAL INSERTS SZ 3, 13MM-1871888
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: 204-23-13 ps tibial inserts sz 3, 13mm-1871888.Device discarded by the customer.Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Event or Problem Description
It was reported a (b)(6) male, initially implanted (b)(6) 2018 on the left side, had a revision approximately 3 years 10 months post initial implant, to exchange the patella and poly due to wear.The patient was revised to exactech devices.Patient was last known to be in stable condition following the event.The explanted devices were discarded by the facility and are not available for return.
 
Additional Manufacturer Narrative
H6: the revision reported was likely the result of prosthesis wear consistent with oxidative degeneration of the polyethylene.Contributing factors to the severity of the wear on the tibial insert may have been result of being packaged in a non-conforming vacuum bag for more than five years and instability of the knee.However, this cannot be confirmed as the devices were not available for evaluation.D4: product information,.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THREE PEG PATELLA 35MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
kate jacobson
2320 nw 66th ct.
gainesville, FL 32653
3523782617
MDR Report Key15053025
Report Number1038671-2022-00812
Device Sequence Number5489676
Product Code JWH
UDI-Device Identifier10885862048325
UDI-Public10885862048325
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 12/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/22/2018
Device Model NumberPS TIBIAL INSERTS SZ 3, 13MM-1871888
Device Catalogue Number204-23-13
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/20/2022
Supplement Date Received by Manufacturer12/15/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date12/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/23/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
200-02-35 - THREE PEG PATELLA 35MM - 5310303; SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexMale
Patient Weight100 KG
-
-