• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL TRIFECTA¿ GT VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL TRIFECTA¿ GT VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number TFGT-A
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Codes Dyspnea (1816); Valvular Insufficiency/ Regurgitation (4449)
Date of Event 10/01/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2018, a 19mm trifecta gt valve was implanted.In (b)(6) 2021, the patient had follow up with no problems reported.In (b)(6) 2021, the patient was admitted for heart failure.In (b)(6) 2022, the patient was admitted again and echocardiography showed severe aortic regurgitation (ar).Medication was administered, however "heart failure did not improve." in (b)(6) 2022, the patient was admitted for additional treatment (oxygen inhalation, medication).Patient symptoms include dyspnea and dizziness.Open-heart surgery was suggested, however the "patient was reluctant." tavr was also considered, however no treatment is scheduled.Conservative therapies was reported after discussion with the patient.The valve remains implanted.No patient consequences were reported.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
An event of regurgitation suspected to be due a leaflet tear was reported.A more comprehensive assessment, including histopathological examination of the valve tissue could not be performed as the device remains implanted and was not returned for analysis.However, based on the information received, the reported leaflet tear is consistent with structural valve deterioration (svd), specifically a non-calcific leaflet tear, which is a well-known complication of tissue heart valve replacement surgery.A variety of factors may contribute to svd, including patient, biological, and implant related factors.No implant related factors could be confirmed from the information received from the field as information related to implant procedure was not provided.Biological factors which can result in tissue degeneration such as calcification (from patient conditions predisposing to elevated calcium like renal disease etc.) or thinning of the prosthetic leaflet tissue (predisposing to leaflet tears) also could not be confirmed as the valve was not returned for histopathological examination.The patient's medical history was not able to be provided.The device history record was also reviewed to ensure that each manufacturing and inspection operation was performed, and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.B3 - date of event is estimated to the first hospital admission due to heart failure.Only the month/year.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIFECTA¿ GT VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15053342
Report Number2135147-2022-00534
Device Sequence Number4821772
Product Code LWR
UDI-Device Identifier05415067018205
UDI-Public05415067018205
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P100029
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2021
Device Model NumberTFGT-A
Device Catalogue NumberTFGT-19A
Device Lot Number6186192
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/22/2022
Supplement Date Received by Manufacturer07/19/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date07/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age82 YR
Patient SexFemale
-
-