| Model Number |
MC2-5590S |
| Medical Device Problem Code |
Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that a revision procedure occurred due to the end cap popping off and the rod no longer working.No patient injury has been reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been returned and is pending evaluation.The root cause cannot be confirmed at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual inspection of the returned product confirmed the threaded cap was unthreaded from the housing tube and the reported failure mode was therefore confirmed.Further inspection showed the returned rod was partially distracted.The laser mark on the nail was confirmed.The issue recorded in this complaint is a previously reported event and had been further investigated as a corrective action was initiated.The root cause is due to variation in the cap tightening process; being unable to verify torque application.Device records review: review of the device history records (dhr) indicates the rod met all required quality inspections and specifications prior to product departure.In addition, the dhr indicates the device was manufactured prior to the latest updated design change.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|