• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ETRINSA 8 DR-T; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ETRINSA 8 DR-T; PACEMAKER Back to Search Results
Model Number 394931
Medical Device Problem Codes Premature Elective Replacement Indicator (1483); Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/18/2022
Type of Reportable Event Malfunction
Event or Problem Description
Error message upon interrogation, battery error.Device was found to be eri.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
The device was not returned for analysis.The analysis is therefore only based on the returned device data as well as the inspection of the quality documents associated with the manufacture of this particular device.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing.The returned device data were inspected.During the inspection the clinical observation could be confirmed.Further analysis revealed an unexpected battery behavior, which led to the clinical observation.Based on the data available for analysis the root cause for that behavior was not determinable and can only be clarified by an analysis of the device itself.However, it cannot be excluded that this occurrence resulted from a defective component, which may compromise the ability of the device to deliver therapy.Therefore, and in order to exclude any potential harm for the patient we would like to recommend replacing the device and to return it for analysis to biotronik.Of course, individual circumstances and medical judgment determine decisions about patient care and the frequency of follow-up sessions.The above recommendation is based on a purely technical assessment.Should further relevant information or the device itself become available this investigation will be updated.
 
Additional Manufacturer Narrative
Device explanted (b)(6) 2022.The device was not returned for analysis.The analysis is therefore only based on the returned device data as well as the inspection of the quality documents associated with the manufacture of this particular device.The manufacturing process for this device was re-investigated.All production steps had been performed accordingly.There was no sign of any inconsistency during the manufacturing.The returned device data were inspected.During the inspection the clinical observation could be confirmed.Further analysis revealed an unexpected battery behavior, which led to the clinical observation.Based on the data available for analysis the root cause for that behavior was not determinable and can only be clarified by an analysis of the device itself.However, it cannot be excluded that this occurrence resulted from a defective component, which may compromise the ability of the device to deliver therapy.Therefore, and in order to exclude any potential harm for the patient we would like to recommend replacing the device and to return it for analysis to biotronik.Of course, individual circumstances and medical judgment determine decisions about patient care and the frequency of follow-up sessions.The above recommendation is based on a purely technical assessment.Should further relevant information or the device itself become available this investigation will be updated.
 
Additional Manufacturer Narrative
Device explanted: (b)(6) 2022.The device was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated and all production steps were performed accordingly.Particularly the final acceptance test proved the device functions to be as specified.Upon initial interrogation a warning message was triggered regarding the battery condition, confirming the clinical observation.The memory content of the device was inspected.The inspection revealed that the battery voltage decreased faster than expected.However, the current consumption was normal.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed in eri mode.There was no indication of a device malfunction.During subsequent analysis the device was opened and subjected to further investigation.The visual inspection of the inner assembly showed no anomalies.The measurement of the battery voltage confirmed an almost depleted battery.A thorough investigation of the electronic module did not show any abnormality.In particular, the current consumption was directly measured and proved to be normal and expected.The battery was sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process, associated with this battery.The battery was subjected to a visual and an electrical inspection.The visual inspection of the battery did not reveal any external signs of damage.The measurement of the battery voltage confirmed an almost depleted battery.Next, the battery was opened for destructive analysis and the inner assembly was inspected.During analysis no internal electrical short was evident.However, the occurrence of a temporary short circuit that contributed to an increased internal self-depletion cannot be excluded.In summary, the root cause for the premature battery depletion could not be determined conclusively.Based on the analysis all therapy functions of the device were available while the device was implanted and in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ETRINSA 8 DR-T
Common Device Name
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key15056722
Report Number1028232-2022-03646
Device Sequence Number11955225
Product Code NVZ
UDI-Device Identifier04035479131821
UDI-Public(01)04035479131821(17)180331
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 07/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2018
Device Model Number394931
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/18/2022
Supplement Date Received by Manufacturer07/20/2022
07/29/2022
01/09/2023
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date07/20/2022
08/02/2022
01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age74 YR
Patient SexMale
-
-