MAUDE Adverse Event Report: EIT EMERGING IMPLANT TECHNOLOGIES GMBH IMPLANT INSERTER SH CONNECTION; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EIT EMERGING IMPLANT TECHNOLOGIES GMBH IMPLANT INSERTER SH CONNECTION; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR Back to Search Results
Model Number AET30101
Medical Device Problem Codes Device-Device Incompatibility (2919); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Event occurred at an unknown date.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from the united kingdom reports an event as follows: it was reported that on an unknown date, during an unknown procedure, the implant inserter in question got stuck on the cage being implanted and would not come apart.The cage was able to be removed from the inserter, but the end of the device was cross-threaded.It was ultimately decided that the cage should not be implanted.There is no reported adverse patient outcome.No further information is available.This report is for an implant inserter sh connection.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANT INSERTER SH CONNECTION
Common Device Name
INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
Manufacturer (Section D)
EIT EMERGING IMPLANT TECHNOLOGIES GMBH
eisenbahnstrasse 84
wurmlingen D-785 73
GM  D-78573
Manufacturer (Section G)
EIT EMERGING IMPLANT TECHNOLOGIES GMBH
DE  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key15058272
Report Number3013730328-2022-00050
Device Sequence Number20039400
Product Code MAX
UDI-Device Identifier04260557774786
UDI-Public(01)04260557774786
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K170503
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAET30101
Device Catalogue NumberAET30101
Device Lot NumberE190DI1200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/21/2022
Initial Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/05/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-