• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERLOC SAFETY INFUSION SET 19G X 0.75IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERLOC SAFETY INFUSION SET 19G X 0.75IN WITH Y-SITE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Medical Device Problem Codes Crack (1135); Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of asfxf043 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported by nurse ¿on (b)(6) 2022, patient had the device connected to them for chemotherapy and the chemotherapy ended up leaking all over the chest wall, nurse cleaned up the chest wall and reported this issue to their physician and the site was then assessed the site, the patient received the rest of their dose.".
 
Additional Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a cracked y-site luer hub was confirmed and the cause was supplier-related.The product returned for evaluation was one 19ga x 0.75¿ powerloc safety infusion set with y-site.Usage residues were observed throughout the sample and needleless injection caps were attached to both luer adapters.A crack was observed in the y-site, opposite the molding gate.The crack appeared to originate at the luer orifice and propagated along a molding weld line in the material.Microscopic inspection of the y-site confirmed that the crack propagated along a weld line feature in the material.The crack exhibited a granular fracture surface.Abundant superficial stress fracturing was observed throughout the region of the y-site near the orifice.The crack occurred along a molding feature in the y-site material.The abundant stress fracturing was also indicative of material embrittlement, likely caused by manufacturing conditions.The device is a supplied component and the supplier has been notified of this event.H3 other text: evaluation findings are in section h11.
 
Event or Problem Description
It was reported by nurse ¿on (b)(6) 2022, patient had the device connected to them for chemotherapy and the chemotherapy ended up leaking all over the chest wall, nurse cleaned up the chest wall and reported this issue to their physician and the site was then assessed the site, the patient received the rest of their dose.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC SAFETY INFUSION SET 19G X 0.75IN WITH Y-SITE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer Contact
kayla olsen
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key15059097
Report Number3006260740-2022-02774
Device Sequence Number20039285
Product Code FPA
UDI-Device Identifier00801741047916
UDI-Public(01)00801741047916
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K060812
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number0671934
Device Lot NumberASFXF043
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/01/2022
Supplement Date Received by Manufacturer09/22/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date09/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/01/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-