INTEGRA LIFESCIENCES MANSFIELD CERELINK ICP PROBE 1L P BOLT; ICP MICRSOSENSORS - CERELINK Back to Search Results
Model Number 826852
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event or Problem Description
A facility reported a cerelink microsensor was implanted and connected to a cerelink monitor.The cerelink monitor started displaying negative values that were not compatible with the patient status (the patient was fine).The value with cerelink monitor was -9.Due to the not reliable values, they decided to measure icp with dve and with dve the icp value was 6.
 
Event or Problem Description
N/a.
 
Additional Manufacturer Narrative
Cerelink sensor was not returned for evaluation (as per customer, product not available) and lot number information has not been provided; therefore, an evaluation of the device could not be performed, and manufacturing records could not be reviewed.The cause(s) of the difficulty reported by the customer could not be determined, however, the possible root cause for the issue reported by the customer could be due to excessive pressure applied to product.If additional relevant information becomes available in the future, this complaint will be reopened, and the respective evaluation performed.Trends will be monitored for this and similar issues.At present, we consider this complaint to be closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERELINK ICP PROBE 1L P BOLT
Common Device Name
ICP MICRSOSENSORS - CERELINK
Manufacturer (Section D)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
11 cabot boulevard
mansfield MA
Manufacturer (Section G)
INTEGRA LIFESCIENCES MANSFIELD
11 cabot boulevard
mansfield MA
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key15059148
Report Number3013886523-2022-00327
Device Sequence Number11955255
Product Code GWM
UDI-Device Identifier10886704075455
UDI-Public10886704075455
Combination Product (Y/N)N
PMA/510(K) Number
K173192
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number826852
Device Catalogue Number826852
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/28/2022
Supplement Date Received by Manufacturer07/29/2022
Initial Report FDA Received Date07/19/2022
Supplement Report FDA Received Date08/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-