Model Number 1005357H
Medical Device Problem Codes Difficult to Remove (1528); Difficult to Advance (2920)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/30/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that the reported kink on the monorail lumen of the stent delivery system (sds) resulted in the reported difficult to advance and the reported difficult to remove.There is no indication of a product quality issue with respect to manufacture, design or labeling.The 2.50x18mm xience skypoint stent referenced is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a 75% stenosed de novo lesion in the distal left circumflex (dlcx) with mild calcification and moderate tortuosity.A bmw guidewire (gw) was positioned and a trek balloon was used for lesion pre-dilatation.Then, the 014 whisper gw was positioned.The 2.50x18mm xience skypoint stent delivery system (sds) was advanced but failed to reach the lesion as it became stuck and caught with the whisper gw.The gw and sds were removed together from the anatomy.A kink on the monorail lumen of the sds was noted.There was no adverse patient effect and no clinically significant delay reported in the procedure.A 2.25mm xience skypoint stent and a non-abbott gw were used to complete the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE WHISPER MS GUIDE WIRE
Common Device Name
GUIDE WIRE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005737652
road no.2 km 58.0 cruce dávila
barceloneta PR 00617
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15059982
Report Number2024168-2022-07915
Device Sequence Number5711752
Product Code DQX
UDI-Device Identifier08717648037801
UDI-Public08717648037801
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model Number1005357H
Device Catalogue Number1005357H
Device Lot Number1091474
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2022
Initial Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.50X18MM XIENCE SKYPOINT STENT
Patient SexUnknown
-
-