| Model Number |
168-420 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
06/20/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Dhr trend summary: no patterns were found; therefore, no additional actions are deemed required.The trend of a050401 - fluid leak complaints will continue to be monitored and corrective actions will be taken in the future if deemed appropriate.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: device deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: ¿ deflation: deflation: observed, one opening device on the posterior side assessed as unidentified (tear) opening (shell thickness within specification).Additional observations: ¿ crease fold observed on the device.¿ yellow encapsulation observed inside device.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side deflation.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional, changed, and/or corrected data: b5, d6b, h.6.
|
| |
|