• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CD2257-40Q FORTIFY ASSURA DR ICD_IDE_US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) CD2257-40Q FORTIFY ASSURA DR ICD_IDE_US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2257-40Q
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Event or Problem Description
Related manufacturer reference number: 2017865-2022-15606.It was reported that the patient presented to the clinic for a device system extraction due to cardiac implantable electronic device (cied) - related endocarditis.The implantable cardioverter defibrillator (icd) and right ventricular lead were explanted.The patient was in stable condition throughout the procedure.
 
Additional Manufacturer Narrative
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CD2257-40Q FORTIFY ASSURA DR ICD_IDE_US
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key15061589
Report Number2017865-2022-15605
Device Sequence Number5529927
Product Code NVZ
UDI-Device Identifier05414734505550
UDI-Public05414734505550
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P910023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2015
Device Model NumberCD2257-40Q
Device Catalogue NumberCD2257-40Q
Device Lot Number4201069
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/13/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/08/2022
Initial Report FDA Received Date07/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/13/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
MEDTRONIC RIGHT ATRIAL LEAD.; MEDTRONIC RIGHT VENTRICULAR LEAD.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
Patient SexMale
Patient Weight115 KG
-
-