|
Medtronic received information that occluded, no delivery/occlusion alarm - unknown timing, time/clock anomaly, exposed to moisture occurred.There was no adverse impact or consequence reported as a result of this event.
|
|
(b)(4).S/w 4.10c.Unit passed the full functional test including the displacement test, rewind, prime/seating test, basic occlusion test and force sensor test.Sleep current measurement and active current measurement within specifications.Pump programmed with the current time and date then monitored for several days.The time and date functioning properly and no time loss/gain anomaly was noted.Unit received with cracked retainer.No moisture damage on electronic assembly, motor and vibrator motor assembly noted.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|