Model Number 183642
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Date of Event 01/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Medical product: tinbn vg int ps anat fm l 70mm: catalog#18132tnbn, lot#3560859; tnbn pol cocr fin tib tray 71: catalog#141253tnbn, lot#2017051899; unknown tibial stem: catalog#ni, lot#ni; unknown patella component: catalog#ni, lot#ni; refobacin bone cement: catalog#ni, lot#ni.Report source: foreign: italy.Customer has indicated that the product will not be returning to zimmer biomet for evaluation, as the product location is unknown.The investigation is in process.Once the investigation has been completed, a follow up report will be submitted.
 
Event or Problem Description
It was reported via a clinical study that a patient underwent an initial left total knee arthroplasty.Approximately one year and two months post implantation, the patient underwent revision surgery due to pain.Due diligence is in progress for this complaint; to date no additional information or product has been received.
 
Event or Problem Description
No further event information at time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections have been updated: visual and dimensional evaluations could not be performed as no product was returned nor were pictures provided.Device history record was reviewed and no discrepancies related to the reported event were found.Root cause was unable to be determined.Per package insert, pain is known adverse effect of the system.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VNGD PS TIB BRG 12X71/75MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15063934
Report Number0001825034-2022-01649
Device Sequence Number5722771
Product Code JWH
UDI-Device Identifier00880304271944
UDI-Public(01)00880304271944(17)220726(10)430510
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K171054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Study,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/26/2022
Device Model Number183642
Device Catalogue Number183642
Device Lot Number430510
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/24/2022
Supplement Date Received by Manufacturer11/08/2022
Initial Report FDA Received Date07/20/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/26/2017
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age66 YR
Patient SexFemale
Patient Weight80 KG
-
-