| Model Number |
N/A |
| Medical Device Problem Code |
Packaging Problem (3007)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/10/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination of the returned product identified the inner blister was not completely sealed as the adhesive residue is not uniform and does not meet the minimum requirements.The device history record was reviewed and no discrepancies relevant to the reported event were found.The likely condition of the product when it left zimmer biomet is non-conforming.However, a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that when the circulator was opening the outer sealed package she stated it came off way too easily.It was determined that in order to be safe she would not pass it to the surgical field.She proceeded to open and pass the inner sealed package.There was no consequences or impact to the patient.
|
| |
|