Model Number 02-022-44-3017
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/23/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Pending evaluation.
 
Event or Problem Description
As reported, approximately 8 years post op the initial left tka, this 67 y/o female patient was revised due to poly wear.Patient was last known to be in stable condition following the event.Device will not be returning.
 
Additional Manufacturer Narrative
Section h10: (h3) the revision reported was likely the result of mechanical overload on the medial side of the joint and subsequent increased wear of the tibial insert due to contributions from instability and/or patient conditions.However, this cannot be confirmed as the device was not available for evaluation and limited clinical information was provided.Section h11: *the following sections have corrected information: (d4) catalog number: 02-012-35-3015, serial number: (b)(6) expiration date: 09-dec-2018, unique identifier (udi) #: (b)(4).(h4) device manufacture date: 10-dec-2010.(h6) component code: 734, bearings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRULIANT TIB IMP PSC INSERT SZ 3, 17MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 ct
gainesville FL 32653
Manufacturer Contact
kate jacobson
3523771140
MDR Report Key15064260
Report Number1038671-2022-00817
Device Sequence Number5628175
Product Code JWH
UDI-Device Identifier10885862001832
UDI-Public10885862001832
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K152170
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/09/2018
Device Model Number02-022-44-3017
Device Catalogue Number02-012-35-3015
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/28/2022
Supplement Date Received by Manufacturer12/01/2022
Initial Report FDA Received Date07/20/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/10/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient SexFemale
-
-