Model Number N/A
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Hematoma (1884); Failure of Implant (1924)
Date of Event 06/23/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient underwent a right knee arthroplasty revision to address a hematoma approximately three (3) months post-operatively.Upon removal, countless vertical lines were found on the surface of the articular surface.Attempts have been made, however, no additional information is available.
 
Additional Manufacturer Narrative
(b)(4).Concomitant devices - persona posterior stabilized standard femoral component right size 9 catalog #: 42500606602 lot #: 65196339, persona cemented stemmed tibial component right size g catalog #: 42532007902 lot #: 65285828.Report source - foreign: japan.The complainant has indicated that the product will not be returned to zimmer biomet for investigation, as the device was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The reported event was able to be confirmed by examination of the provided pictures.Inspection of the pictures identified pitting and wear lines on the implant.Review of the device history records identified no deviations or anomalies during manufacturing.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSONA POSTERIOR STABILIZED ARTICULAR SURFACE RIGHT 10MM
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
Manufacturer (Section G)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI  
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key15064292
Report Number3007963827-2022-00187
Device Sequence Number5722779
Product Code MBH
UDI-Device Identifier00889024243910
UDI-Public(01)00889024243910(17)260803(10)65004588
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K121771
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/10/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/03/2026
Device Model NumberN/A
Device Catalogue Number42522400910
Device Lot Number65004588
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer11/01/2022
Initial Report FDA Received Date07/20/2022
Supplement Report FDA Received Date11/10/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/30/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10 NARRATIVE
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight74 KG
-
-