• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOTAK RELIANCE G; IMPLANTABLE LEAD Back to Search Results
Model Number 0262
Medical Device Problem Codes Defibrillation/Stimulation Problem (1573); Failure to Read Input Signal (1581); Under-Sensing (1661)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) exhibited low amplitude and undersensing on the right ventricular (rv) lead.Which led to a delay in the therapy.Subsequently, the rv lead was not successfully implanted.This rv lead was never in service and no adverse patient effects were reported.
 
Additional Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, analysis of the returned lead found no evidence of device defect, malfunction, or damage outside the bounds of normal medical use.The device was subjected to and passed all automated testing.Laboratory analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Event or Problem Description
It was reported that the patient with this implantable cardioverter defibrillator (icd) exhibited low amplitude and undersensing on the right ventricular (rv) lead.Which led to a delay in the therapy.Subsequently, the rv lead was not successfully implanted.This rv lead was never in service and no adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOTAK RELIANCE G
Common Device Name
IMPLANTABLE LEAD
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15064453
Report Number2124215-2022-26485
Device Sequence Number5504161
Product Code LWS
UDI-Device Identifier00802526601828
UDI-Public00802526601828
Combination Product (Y/N)Y
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P910073/S077
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date02/10/2024
Device Model Number0262
Device Catalogue Number0262
Device Lot Number100487
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer09/07/2022
Initial Report FDA Received Date07/20/2022
Supplement Report FDA Received Date09/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/10/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age44 YR
Patient SexFemale
-
-