The details provided about the alleged incident were the following: the physician had attempted to advance a placed a 6mm x 22mm x 120cm long icast covered stent through a 110cm sheath yet it didn't come out on the other side.He removed the stent, opened another 120cm icast & laid them next to each other on a table, and one was 5-10cm shorter.The device in question was returned for evaluation.The usable length of the catheter was measured from the end of the manifold strain relief to the distal tip of the catheter, inclusive of the length of the balloon.The usable length of the catheter was 117.5cm.The product specification for the catheter usable length is 120cm +/- 5cm per part specification drawing ps010634-xxx ps, balloon catheter, otw stent delivery, 3l, 6mm revision af.The returned product meets this requirement.Useable length of the catheter is measured for every device during manufacturing per mp009165, manufacturing final inspection, balloon catheter, and otw stent delivery, to ensure the length meets specification after the catheter is assembled.The balloon segment of the catheter that has the stent crimped to it was also measured from the proximal balloon weld to the distal tip of the catheter.This measured 4.5cm.This measurement is of the working part of the device that must be clear of the introducer sheath prior to deployment.Based on the usable length of the catheter being 117.5cm and the balloon segment measuring 4.5cm, the balloon would have cleared the entire length of a 110cm introducer sheath by 3cm.The actual length of the introducer sheath is not known and was not returned for evaluation.Images of the other device that was claimed to have been longer than the returned device were also not provided.Based on the details of the complaint and the results of the investigation the complaint cannot be confirmed as the product when measured met the usable length requirements as specified above.There is no indication that the device in question was non-conforming or faulty.No nonconformance were identified in the device history review.A trend review was conducted and this has been the only complaint listed for the reported defect of ¿incompatible accessory used during procedure/ prolonged procedure, no intervention required" in the last 15 months.No capa requests or capas or ncrs were found to be related to this event.A risk review found that the risk management documents for this product adequately address the reported defect and the severity and anticipated occurrence level are appropriate.Labeling was found to have appropriate information pertaining to device sizing and use of accessories; however, the device in this case was used off-label.
|