Model Number 20893-000
Medical Device Problem Code Radiation Leak (1357)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/22/2022
Type of Reportable Event Malfunction
Event or Problem Description
During an ablation procedure, it was reported that the user noticed the probe was not heating while firing the laser.The clinical specialist went into the mri room to check the connections and discovered the probe's laser fiber was damaged and was warm to the touch at the location of the damage.Upon discovery of the damage, the laser light was shining through damaged area of cable.There were no patient complications reported in relation to this event.
 
Additional Manufacturer Narrative
The neuroblate system instructions for use contain laser safety warnings as risk mitigations for this event: section 4.3, general warnings: laser eye protection provided with the neuroblate system must be worn in the mri scanner room during operation of the laser.To avoid equipment damage and patient injury, do not continue to deliver laser energy if no elevation in tissue temperature is observed.To avoid equipment damage and patient injury, do not continue to deliver laser energy if no elevation in tissue temperature is observed.No harm was observed in this event or any previously reported event, and the event description provided by the clinical specialist indicates that these instructions for risk mitigation were followed.Therefore, these risk mitigations remain effective.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROBLATE OPTIC FULLFIRE® LASER PROBE
Common Device Name
NEUROBLATE OPTIC FULLFIRE® LASER PROBE
Manufacturer (Section D)
MONTERIS MEDICAL
131 cheshire ln
suite 100
minnetonka MN 55305
Manufacturer (Section G)
MONTERIS MEDICAL
131 cheshire ln
suite 100
minnetonka MN 55305
Manufacturer Contact
jordon honeck
131 cheshire ln
suite 100
minnetonka, MN 55305
MDR Report Key15064567
Report Number3009970070-2022-00019
Device Sequence Number12144707
Product Code GEX
UDI-Device Identifier00816589020953
UDI-Public00816589020953
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
K182036
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/23/2023
Device Model Number20893-000
Device Catalogue NumberDTP-000-01
Device Lot Number11333
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/22/2022
Initial Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-