| Model Number |
N/A |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This event has been recorded by zimmer biomet under (b)(4).Investigation is complete.This is an initial final report submission.Review of the most recent repair record determined the cart would not power on and the power inlet module was burnt.The power inlet module was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
It was reported that prior to surgery this cart would not power up.There was no patient involvement.At product evaluation investigation power inlet module was burnt.No adverse events were reported as a result of this malfunction.No additional event information available.
|
| |
|
Search Alerts/Recalls
|