• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE Back to Search Results
Model Number 777726
Medical Device Problem Codes Material Fragmentation (1261); Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/04/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that jj tip of the ureteral stent was split.Another device was used.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Additional Manufacturer Narrative
The reported event is inconclusive because no sample was returned for evaluation.A potential root cause for this event could be, "material selection".The device was used for treatment purposes.It is unknown if the device had met all relevant specifications or resulted in the reported event.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "improper handling technique can seriously weaken the stent.Acute bending or overstressing during placement could result in subsequent separation of the stent at the point of stress after a prolonged indwelling period." "exercise care.Tearing of the stent can be caused by sharp instruments." "care should be exercised when removing the stent from inner polybag so as not to cause tearing or fragmentation." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that jj tip of the ureteral stent was split.Another device was used.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® INLAY® URETERAL STENT WITH HYDROGLIDE¿ GUIDEWIRE
Common Device Name
GUIDEWIRE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
juan velez
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15064857
Report Number1018233-2022-05632
Device Sequence Number5425652
Product Code FAD
UDI-Device Identifier10801741014625
UDI-Public(01)10801741014625
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K983498
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/03/2026
Device Model Number777726
Device Catalogue Number777726
Device Lot NumberNGFZ4152
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/04/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date07/20/2022
Supplement Report FDA Received Date12/12/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/26/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-