| Model Number |
1581-22-010 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Off-Label Use (1494)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/15/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Upon implantation, tibial insert was not completely flush laterally leaving space between the tibial tray and the insert.Operated knee is right.Surgeon decided to change insert size to 10mm but same issue was observed.Final decision was to implant size 10 even with the problem involved.Fear of insert dislocation and backside wear remains.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # = > pc-(b)(4).Investigation summary = > no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = > a manufacturing record evaluation (nc search) was performed for the finished device 158122010, lot d21090791, and no non-conformances / manufacturing irregularities were identified.
|
| |
|
Search Alerts/Recalls
|