| Model Number |
358764 |
| Medical Device Problem Codes |
Failure to Capture (1081); Fracture (1260); High impedance (1291)
|
| Health Effect - Clinical Code |
Asystole (4442)
|
| Date of Event |
06/07/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Boston scientific provided the following information: this rv lead was capped due to noise, oversensing, loss of capture and asystole greater than 2 seconds.A lead fracture and impedances greater than 2500 ohms were observed as well.
|
| |
|
Additional Manufacturer Narrative
|
|
The device is currently not available for analysis.No conclusion can be drawn at this time.No additional information is available at the moment.The file is closed.The investigation will be re-opened should additional data become available.
|
| |
|