| Model Number |
359073 |
| Medical Device Problem Code |
Over-Sensing (1438)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/14/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
After an implantation period of approximately 142 months it was reported that the lead was capped due to oversensing in the ventricular channel.
|
| |
|
Additional Manufacturer Narrative
|
|
The lead under complaint was not returned for analysis.This report is therefore only based on the analysis of the icd itself as well as the inspection of the quality documents associated with the manufacture of the lead and the icd.The manufacturing process for the devices was re-investigated, revealing that all production steps were performed accordingly.There was no sign of any inconsistency during the manufacturing process.Particularly the final acceptance test proved the devices functions to be as specified.Subsequently the device was interrogated.The interrogation could be properly performed and revealed the mos1 battery status.All electrical parameters were verified and showed a normal device behavior.Especially, the current consumption and the battery state of discharge were found normal and as expected.There was no indication of a device malfunction.
|
| |
|