Manufacturer's investigation conclusion: incidental findings: atypical power cable disconnect alarms activated due to black power cable faults activating.The reported event of backup battery fault alarms was confirmed via analysis of the submitted log file.The log file contained approximately 6 days of data (22jun2022 ¿ 28jun2022 per the timestamp).Multiple backup battery fault alarms activated on 22jun2022 from 10:43:11 to 10:43:12 and on 27jun2022 from 21:09:12 to 21:24:37 due to backup battery circuit faults activating; the alarms resolved shortly after activating each time.The pump maintained a speed above the low speed limit throughout the log file.The heartmate 3 system controller serial number: (b)(4).Was not returned for evaluation.Technical services recommended reseating the backup battery in the controller and exchanging the backup battery if the issue is not resolved after reseating the product.Multiple requests for additional information were made to determine if the backup battery fault alarms resolved, if any products were exchanged and if any products would be returned for evaluation; however, no response was received.The root cause of the reported event could not be conclusively determined through this analysis.Heartmate 3 patient handbook (rev.D) section 5, entitled ¿alarms and troubleshooting¿, and heartmate 3 instructions for use (ifu) (rev.C) section 7, entitled ¿alarms and troubleshooting¿, cover all alarms (visual and audible), including the backup battery fault and power cable disconnect alarm conditions, and the actions to take if the alarms cannot be resolved.Heartmate 3 instructions for use (rev.C) section 2, entitled "system operations", explains how to replace the system controller and how to replace the system controller backup battery.Section 5, entitled ¿surgical procedures¿, also explains how to install the backup battery in the system controller.Heartmate 3 patient handbook (rev.D) section 6, entitled "caring for the equipment", describes how to care for and clean all equipment, including the 14v battery clips and 14v batteries.Section 10, entitled ¿safety checklists¿, provides checklists to assist the patient in performing routine maintenance of heartmate 3 lvad, including inspecting/cleaning the contacts on the 14v battery clips and 14v batteries.The heartmate 3 patient handbook (rev.C) cautions the user to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.The device history records were reviewed and the records revealed that the heartmate 3 system controller, serial number (b)(4), was manufactured in accordance with manufacturing and qa specifications.The heartmate 3 system controller was shipped to the customer on (b)(6) 2022.No further information was provided.The manufacturer is closing the file on this event.
|