| Model Number |
TRM-275 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.The reason for reoperation is: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare provider reported a right side capsular contracture baker grade iv.The device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: "visual analysis of the returned device identified deformation.A weight test of the device was verified and the device was within specification.Cloudy after autoclave disinfection process in the gel.Based on the device analysis the final assessment is: no issues found related with the manufacturing.".
|
| |
|
Event or Problem Description
|
|
Patient reported a right side capsular contracture baker grade iv.The device was explanted.
|
| |
|
Search Alerts/Recalls
|