U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 5MM X 15CM; DEVICE, NEUROVASCULAR EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER NEUROVASCULAR CORK TARGET XL 360 SOFT 5MM X 15CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0036125150
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not available to the manufacturer.
 
Event or Problem Description
It was reported that during procedure the coil (subject device) broke off inside the catheter.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Event or Problem Description
It was reported that during procedure the coil (subject device) broke off inside the catheter.The procedure was completed successfully.No clinical consequences were reported to the patient due to this event.
 
Additional Manufacturer Narrative
Section b1 adverse event/product problem - corrected - no product problem.Section h1 type of reportable event - corrected - no malfunction.H4 manufacturing date ¿ added.H3 device evaluated by mfg ¿updated.H3 summary attached - updated.D4 expiration date - added.D9 product available to stryker ¿ updated.D9 returned to manufacturer on ¿updated.Due to the automated mes (manufacturing execution system) system, there are controls in the manufacturing process to ensure the product met specifications upon release.During visual inspection, the device was returned stuck in a non-stryker catheter, the main coil was seen to be protruding out of the distal end of the non-stryker catheter, the main coil was seen to be kinked/bent and stretched.No breakage noted, and the coil delivery wire was seen to be kinked/bent.During a functional test, the device was stuck in a non-stryker catheter and, despite efforts made, could not be retracted or pushed through the catheter.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint was not confirmed based on analysis.The device failed to meet specifications when received for complaint investigation based on the analyzed anomalies noted to the device.During analysis, the main coil was found to be stuck inside the microcatheter.The coil was found to be kinked/bent and stretched; however, there was no breakage noted.An assignable cause of procedural factors will be assigned to the analyzed ¿main coil stretched¿, ¿main coil kinked/bent¿, and ¿moil jammed¿ as the defects are associated with a product that met stryker design and manufacturing specifications and was used in accordance with the dfu but performance was limited due to procedural and/or anatomical factors during use.An assignable code of not confirmed will be assigned to the reported 'main coil broken/fractured during use' as the defect was not confirmed during the analysis.An assignable cause of handling damage will be assigned to the analyzed 'coil delivery wire kinked/bent' since it is most likely that this damage occurred due to handling of delivery wire during the clinical procedure.The manufacturer has reviewed all information and determined this event no longer meets the requirement of the reportable event for the device in question.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET XL 360 SOFT 5MM X 15CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15070861
Report Number3008881809-2022-00362
Device Sequence Number5489421
Product Code HCG
UDI-Device Identifier07613327128291
UDI-Public07613327128291
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2025
Device Model NumberM0036125150
Device Catalogue NumberM0036125150
Device Lot Number23474783
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/05/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date07/20/2022
Supplement Report FDA Received Date09/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PHENOM 17 MICROCATHETER (NON-STRYKER)
Patient SexUnknown
-
-