• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOVASCULAR LLC VASOVIEW 7 XB BISECTOR (US); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOVASCULAR LLC VASOVIEW 7 XB BISECTOR (US); ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number VASOVIEW 7 XB BISECTOR (US)
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
The hospital reported that during an endoscopic vein harvesting procedure using vasoview 7 xb bisector, they were taking vein on friday and towards the end of harvesting they were cutting some fascial tissue to open up the tunnel and the blade got stuck on one of the prongs and they were unable to get the knife in or out.There was no damage to the vein.No delay.No parts fell off.Device retrieved today.Used another device to complete case.No harm to patient.
 
Additional Manufacturer Narrative
Trackwise id (b)(4).The device has been returned to the factory and is being evaluated.A supplemental report will be submitted when the evaluation is completed.
 
Additional Manufacturer Narrative
Trackwise # (b)(4).The device was returned to the factory for evaluation on 07/14/2022.An investigation was conducted on 08/03/2022.Both the harvesting device and cannula were returned.A visual inspection was conducted.Signs of clinical use and evidence of blood were observed.The c-ring was observed to be intact with no visual or physical defects.A mechanical evaluation was conducted.The blue toggle was manipulated to attempt to extend and retract the cutter blade.The toggle would not move and the cutter blade would not retract or extend.The cutter blade was observed to stuck in between the bipolar blades at the base of the jaws.No other visual defects were observed.Based on the returned condition of the device and the results of the investigation, the reported failure "mechanical problem" was confirmed.The lot history record review was completed.There were no ncmrs, rework, or deviations documented for the reported lot number.Based on the dhr/lhr review results, it was determined that there is no relation between the batch manufacturing process and the reported failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VASOVIEW 7 XB BISECTOR (US)
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer (Section G)
MAQUET CARDIOVASCULAR LLC
45 barbour pond drive
wayne NJ
Manufacturer Contact
arelean guzman
45 barbour pond drive
wayne, NJ 
MDR Report Key15071033
Report Number2242352-2022-00628
Device Sequence Number14780973
Product Code GEI
UDI-Device Identifier00607567700369
UDI-Public00607567700369
Combination Product (Y/N)N
Initial Reporter StateNY
PMA/510(K) Number
K041981
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician Assistant
Type of Report Initial,Followup
Report Date (Section B) 09/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/10/2004
Device Model NumberVASOVIEW 7 XB BISECTOR (US)
Device Catalogue NumberVH-3200
Device Lot Number3000234333
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/05/2022
Supplement Date Received by Manufacturer09/08/2022
Initial Report FDA Received Date07/20/2022
Supplement Report FDA Received Date09/08/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/11/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.; UNKNOWN.
Patient SexMale
-
-