• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD NEOPUFF INFANT RESUSCITATOR; BTL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD NEOPUFF INFANT RESUSCITATOR; BTL Back to Search Results
Model Number RD900
Medical Device Problem Code Pressure Problem (3012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
A healthcare facility in france reported via a fisher and paykel (f&p) representative that pressure of an rd900 neopuff infant resuscitator unit would not reach 30cm at 8l per minute.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).Method: the complaint rd900 neopuff infant resuscitator was returned to the fisher & paykel healthcare (f&p) service center in france, where it was inspected, and performance tested by a trained f&p service technician.Results: inspection and performance testing of the rd900 neopuff infant resuscitator revealed that there was no fault.Conclusion: we are unable to determine the cause of the reported event.The neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infants until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to impact damage, for instance when accidentally dropped or subjected to considerable external force.The neopuff technical manual warns against dropping the neopuff or subjecting it to impact damage which may cause the unit to operate incorrectly.If the neopuff is suspected to have been damaged, the manometer and valve system should be performance tested.Each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications.Any unit that fails is rejected.The subject neopuff would have met the requirements at the time of production.The neopuff technical manual states the following: dropping the neopuff / perivent infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient.In addition, the neopuff user instructions state that the user should "check manometer reads zero with no gas flow" and check the pressure settings "prior to every use of the neopuff".Each neopuff unit is assembled and 100% tested in the production line to verify that each unit conforms to critical product specifications.Any unit that fails is rejected.The subject neopuff would have met the required specification at time of production.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEOPUFF INFANT RESUSCITATOR
Common Device Name
BTL
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534002
MDR Report Key15071119
Report Number9611451-2022-00652
Device Sequence Number5339023
Product Code BTL
UDI-Device Identifier09420012410948
UDI-Public(01)09420012410948(10)2100550805(11)180806
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K892885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberRD900
Device Catalogue NumberRD900
Device Lot Number2100550805
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/01/2022
Initial Report FDA Received Date07/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/06/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-