• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INJECTION GOLD PROBE; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INJECTION GOLD PROBE; UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number M00560150
Device Problems Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/24/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the problem analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that an injection gold probe needle was used during hemostasis procedure performed on (b)(6) 2022.During the preparation, it was found that injection needle is not coming out from the catheter.The procedure was not completed successfully.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Block h6: impact code f1001 is being used to capture the reportable issue of aborted/cancelled procedure.Block h10: investigation results: the returned optiflo hemostasis catheter was analyzed, and a visual evaluation noted that no damages or additional defects.A functional evaluation noted that the needle did not extend.Additionally, x ray analysis observed that the needle was kinked and broken.Based on all available information and the condition of the returned device, it is most likely procedural factors such as lesion characteristics, handling of the device, the technique used by the physician during the manipulation of the handle, and normal procedural difficulties encountered during the procedure that could have resulted in the damages encountered in the needle.The investigation concluded the most probable cause is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
 
Event Description
It was reported to boston scientific corporation that an injection gold probe needle was used during hemostasis procedure performed on (b)(6) 2022.During the preparation, it was found that injection needle is not coming out from the catheter.The procedure was not completed successfully.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INJECTION GOLD PROBE
Type of Device
UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15071845
MDR Text Key304516174
Report Number3005099803-2022-03743
Device Sequence Number1
Product Code KNS
UDI-Device Identifier08714729056492
UDI-Public08714729056492
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K133933
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/21/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/21/2022
Device Model NumberM00560150
Device Catalogue Number6015
Device Lot Number0025748827
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/08/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-