MAUDE Adverse Event Report: MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 2; CEMENTLESS HIP STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDACTA INTERNATIONAL SA STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 2; CEMENTLESS HIP STEM Back to Search Results
Model Number 01.12.032
Health Effect - Clinical Code Inadequate Osseointegration (2646)
Date of Event 06/27/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Batch review performed on 28 june 2022: lot 152229: (b)(4) items manufactured and released on 20-aug-2015.Expiration date: 2020-07-13.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been sold without any similar reported event during the period of review.Clinical evaluation performed by medical affairs department: femoral component revision performed 7 years after primary cementless total hip arthroplasty in 70 year old male patient.No information concerning patient general health status and the presence of comorbidities is available.No real radiographic images provided that does not permit to have an evaluation of the whole construct.In the images provided, radiolucent lines are visible.Aseptic loosening is a possible literature described adverse event after primary cementless hip arthroplasties and causes are often unknown.The reason of this failure cannot be determined.
 
Event or Problem Description
At about 7 years after the primary, revision surgery was performed following aseptic stem loosening.The surgeon revised successfully stem, inlay and head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEM: QUADRA-H CEMENTLESS, HA COATED LAT STEM SIZE 2
Common Device Name
CEMENTLESS HIP STEM
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, 6874
SZ   6874
Manufacturer Contact
stefano baj
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key15072643
Report Number3005180920-2022-00550
Device Sequence Number14780984
Product Code JDI
UDI-Device Identifier07630030802140
UDI-Public07630030802140
Combination Product (Y/N)N
Initial Reporter CountrySZ
PMA/510(K) Number
K082792
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/13/2020
Device Model Number01.12.032
Device Catalogue Number01.12.032
Device Lot Number152229
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2022
Initial Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/20/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient SexMale
Patient Weight80 KG
Patient RaceWhite
-
-