| Catalog Number |
18965050S |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Implant Pain (4561)
|
| Date of Event |
10/08/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.A review of the device history is not possible because the lot number was not communicated.Should additional information become available, it will be provided in a supplemental report.Device disposition unknown.
|
| |
|
Event or Problem Description
|
|
It was reported through the filing of a lawsuit that allegedly on (b)(6) 2021 the patient tripped and fell and on (b)(6) 2021 she presented for treatment and was diagnosed with a right subtrochanteric femur fracture.Allegedly on (b)(6) 2021 she underwent an intramedullary nail fixation and cerclage wiring in the subtrochanteric region utilizing gamma 3.Allegedly on (b)(6) 2021 the patient developed sudden onset of severe pain in the right hip and it was noted the proximal aspect of the rod was fractured at the location of the proximal compression screw and distal interlocking screw was fractured.Revision surgery was performed on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could be confirmed on provided ray.The screw had broken in the middle of the shaft.The item was not available for physical examination as it is subject to a legal case.Without the screw the kind of breakage could not be determined.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.Potential causes of implant breakage are listed in the labelling.According to received medical records, incl.X-rays, revealed that in this case a today 75-year-old female patient was treated with above screw on (b)(6) 2021 due to a displaced subtrochanteric fracture of the right femur.On (b)(6) 2021 a visit report confirms the screw was already broken in the mid of august.An operative report stated that the implants were revised with a competitor¿s nail due to screw and nail breakage on (b)(6) 2021.Received x-rays and medical records will be reviewed separately by a medical professional.From technical point of view a root cause for screw breakage could not be determined.From medical point of view the fracture pattern, initial insufficient compression and suboptimal treatment due to lateral entry point contributed to the event.If the device is returned or if substantial information is provided, the investigation will be reassessed.
|
| |
|
Event or Problem Description
|
|
It was reported through the filing of a lawsuit that allegedly on june 28, 2021 the patient tripped and fell and on (b)(6) 2021 she presented for treatment and was diagnosed with a right subtrochanteric femur fracture.Allegedly on (b)(6) 2021 she underwent an intramedullary nail fixation and cerclage wiring in the subtrochanteric region utilizing gamma 3.Allegedly on (b)(6) 2021 the patient developed sudden onset of severe pain in the right hip and it was noted the proximal aspect of the rod was fractured at the location of the proximal compression screw and distal interlocking screw was fractured.Revision surgery was performed on (b)(6) 2021.
|
| |
|
Search Alerts/Recalls
|