Manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one ultraverse 035 pta catheter was returned for evaluation.No specific anomalies were noted during the visual evaluation.In the functional testing, the device was inflated with an in-house presto inflation device and water was leaking from the pin-hole balloon rupture.No further testing was performed.All the anomalies are noted on microscopic observation.Therefore, the investigation is confirmed for the reported balloon rupture as a pin-hole rupture was observed during the microscopic observation.A definitive root cause for the reported balloon rupture could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.D4 (expiry date: 01/2025).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|