• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY UNI TIBIAL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY UNI TIBIAL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71441353
Medical Device Problem Codes Fracture (1260); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/16/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a journey uni tibial impactor broke.As this was noticed upon field inspection, there was not patient involvement.Investigation also determined that the bumper broke and the dowel pin intended for connecting the bumper to the tibial impactor handle is bent.
 
Additional Manufacturer Narrative
The associated device was returned and evaluated.Upon visual inspection, it was confirmed that the tibial impactor is broken.The bumper part is missing and was not returned.The dowel pin intended for connecting the bumper to the tibial impactor handle is bent.A review of complaint history based on the historical data revealed similar events for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it cannot longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY UNI TIBIAL IMPACTOR
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, NON-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
17000 west william cannon dri
austin, TX 78735
5123913905
MDR Report Key15073425
Report Number1020279-2022-03391
Device Sequence Number5377540
Product Code HSX
UDI-Device Identifier03596010588050
UDI-Public03596010588050
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K113038
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number71441353
Device Catalogue Number71441353
Device Lot Number13LM14592
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2022
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/29/2022
Initial Report FDA Received Date07/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2013
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-