• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH DOBHFF 8FR 43IN W STYLTY SITE; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH DOBHFF 8FR 43IN W STYLTY SITE; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number 8884710859
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event or Problem Description
The customer reported that they inserted dobhoff nasogastric tube and went to remove the guidewire and it was stuck.Cardinal health sales rep visited with intensive care unit (icu) and emergency room (er), the staff stated that they got the wire out after significant trouble.There was no patient harm reported.
 
Additional Manufacturer Narrative
The lot number was provided, and the device history record (dhr) was reviewed indicating that the product was released accomplishing all quality standards.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.The device was manufactured on 11-oct-2019.A sample was not received for the investigation.Because a sample was not returned, we were unable to perform a follow up investigation to include functional and visual evaluations to confirm the issue and root cause analysis.At this time, a corrective and preventive action is not deemed necessary.We will keep monitoring the process for any adverse trends that require immediate attention.This complaint will be used for qa tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DOBHFF 8FR 43IN W STYLTY SITE
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
calle 9 sur no. 125 cuidad ind
tijuana 22500
MX   22500
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key15073868
Report Number9612030-2022-03328
Device Sequence Number14831635
Product Code KNT
UDI-Device Identifier10884521518766
UDI-Public10884521518766
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number8884710859
Device Catalogue Number8884710859
Device Lot Number1928300364
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/15/2022
Supplement Date Received by Manufacturer07/15/2022
Initial Report FDA Received Date07/21/2022
Supplement Report FDA Received Date08/11/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/11/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-